Specification
Rocket Medical PLC are award winning designers and manufacturers of medical devices for a wide range of clinical applications. These medical devices are packed into prepurchased pouches and sealed in-house before being sent offsite for sterilisation. The heat seal on these pouches must conform to British Standard BS EN 868-5:1999 (Packaging materials and systems for medical devices which are to be sterilised). Wishing to perform this test themselves, Rocket Medical sought a test system, which could accurately confirm that the strength of each heat seal joint was not below 1.5 N per 15 mm width before and after undergoing the sterilisation process.
Solution
Mecmesin provided an Imperial 1000N test system (now superseded by the MultiTest-i range), a 250 N loadcell and lever-operated serrated cam grips.To perform the test, 15mm width strips are cut at right angles to the heat seal joint at the mid-point of the pouch. Each end of the strip is then clamped into opposing jaws on the test stand with the tail unsupported. The seal is tested at a rate of 200 mm/min ±10 mm/min and the maximum force recorded. The computer-controlled test system offered Rocket Medical full programming capability. The system displays the maximum force achieved during the test, as well as the average maximum force incurred over a number of tests. This saves the company a significant amount of time when performing tests and provides them with a means of guaranteeing compliance with the relevant standards.
Testimonial
‘In order to meet the British Standard regarding heat seal strengths on medical pouches, Mecmesin provided an operator-friendly system, which displays data in graphical format to clearly show compliance to the standard.’
David Dixon, Quality Assurance Technician
Rocket Medical PLC, UK